Human Interleukin-15 receptor subunit alpha(IL15RA) ELISA kit CSB-EL011594HU CUSABIO
$50.00
In stock
SKU
CSB-EL011594HU
Product Details
| Target Name | interleukin 15 receptor, alpha |
| Uniprot No. | Q13261 |
| Species | Homo sapiens (Human) |
| Storage | Store at 2-8°C. Please refer to protocol. |
| Sensitivity | 7.8 pg/mL |
| Troubleshooting and FAQs | ELISA kit FAQs |
| Lead Time | 3-5 working days after you place the order, and it takes another 3-5 days for delivery via DHL or FedEx. |
| Associated names | AA690181 ELISA Kit; CD215 ELISA Kit; I15RA_HUMAN ELISA Kit; IL 15R alpha ELISA Kit; IL-15 receptor subunit alpha ELISA Kit; IL-15R-alpha ELISA Kit; IL-15RA ELISA Kit; Il15ra ELISA Kit; Interleukin 15 receptor alpha ELISA Kit; Interleukin 15 receptor subunit alpha ELISA Kit; MGC104179 ELISA Kit; sIL-15 receptor subunit alpha ELISA Kit; sIL-15R-alpha ELISA Kit; sIL-15RA ELISA Kit; Soluble interleukin 15 receptor subunit alpha ELISA Kit; Soluble interleukin-15 receptor subunit alpha ELISA Kit |
| Abbreviation | IL15RA |
| Sample Types | serum, plasma, tissue homogenates |
| Detection Range | 31.25 pg/mL-2000 pg/mL |
| Assay Time | 1-5h |
| Sample Loading Volume | 50-100ul |
| Detection Wavelength | 450 nm |
| Research Area | Immunology |
| Assay Principle | quantitative |
| Measurement | Sandwich |
| Precision | headers: Intra-assay Precision (Precision within an assay): CV%<8%; rows: Intra-assay Precision (Precision within an assay): CV%<8%: Intra-assay Precision (Precision within an assay): CV%<8%; 1: Three samples of known concentration were tested twenty times on one plate to assess.; Intra-assay Precision (Precision within an assay): CV%<8%: Inter-assay Precision (Precision between assays): CV%<10%; 1: Three samples of known concentration were tested in twenty assays to assess.; |
| Linearity | headers: To assess the linearity of the assay, samples were spiked with high concentrations of human IL15RA in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.; rows: To assess the linearity of the assay, samples were spiked with high concentrations of human IL15RA in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.: To assess the linearity of the assay, samples were spiked with high concentrations of human IL15RA in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.; 2: Sample; 3: Serum(n=4); 1: 1:1; 2: Average %; 3: 93; 1: Range %; 2: 88-97; 1: 1:2; 2: Average %; 3: 95; 1: Range %; 2: 91-99; 1: 1:4; 2: Average %; 3: 92; 1: Range %; 2: 87-97; 1: 1:8; 2: Average %; 3: 89; 1: Range %; 2: 85-93 |
| Recovery | headers: The recovery of human IL15RA spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.; rows: The recovery of human IL15RA spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.: The recovery of human IL15RA spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.; ; 1: Sample Type; 2: Average % Recovery; 3: Range; 1: Serum (n=5); 2: 91; 3: 88-93; 1: EDTA plasma (n=4); 2: 104; 3: 101-108; ; |
| Typical Data | headers: These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.; rows: These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.: These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.; ; 1: pg/ml OD1 OD2 Average Corrected 2000 2.766 2.844 2.805 2.726 1000 2.243 2.253 2.248 2.169 500 1.348 1.366 1.357 1.278 250 0.771 0.805 0.788 0.709 125 0.352 0.366 0.359 0.280 62.5 0.256 0.277 0.267 0.188 31.25 0.149 0.154 0.152 0.073 0 0.077 0.080 0.079; 2: pg/ml; 3: OD1; 4: OD2; 5: Average; 6: Corrected; 9: 2000; 10: 2.766; 11: 2.844; 12: 2.805; 13: 2.726; 16: 1000; 17: 2.243; 18: 2.253; 19: 2.248; 20: 2.169; 23: 500; 24: 1.348; 25: 1.366; 26: 1.357; 27: 1.278; 30: 250; 31: 0.771; 32: 0.805; 33: 0.788; 34: 0.709; 37: 125; 38: 0.352; 39: 0.366; 40: 0.359; 41: 0.280; 44: 62.5; 45: 0.256; 46: 0.277; 47: 0.267; 48: 0.188; 51: 31.25; 52: 0.149; 53: 0.154; 54: 0.152; 55: 0.073; 58: 0; 59: 0.077; 60: 0.080; 61: 0.079; 1: pg/ml; 2: OD1; 3: OD2; 4: Average; 5: Corrected; 1: 2000; 2: 2.766; 3: 2.844; 4: 2.805; 5: 2.726; 1: 1000; 2: 2.243; 3: 2.253; 4: 2.248; 5: 2.169; 1: 500; 2: 1.348; 3: 1.366; 4: 1.357; 5: 1.278; 1: 250; 2: 0.771; 3: 0.805; 4: 0.788; 5: 0.709; 1: 125; 2: 0.352; 3: 0.366; 4: 0.359; 5: 0.280; 1: 62.5; 2: 0.256; 3: 0.277; 4: 0.267; 5: 0.188; 1: 31.25; 2: 0.149; 3: 0.154; 4: 0.152; 5: 0.073; 1: 0; 2: 0.077; 3: 0.080; 4: 0.079 |
| ELISA Data Analysis | ELISA Standard Curve & Curve Expert software |
| Shelf Life | 6 months |
| Description | This Human IL15RA ELISA Kit was designed for the quantitative measurement of Human IL15RA protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 31.25 pg/mL-2000 pg/mL and the sensitivity is 7.8 pg/mL. |
MDSD