| Product name | Human APOH ELISA Kit |
| Synonyms | Apolipoprotein H, Activated protein C-binding protein, Apo H, Apo-H, ApolipoproteinH |
| Reactivity | Human |
| Host / Isotype | United States |
| Tests | 1 X 96 well plate |
| Sample type | Serum, Plasma, Cell lysate |
| Assay type | Sandwich |
| Sensitivity | 6.2 pg/mL |
| Range | 46.88-3000 pg/mL |
| Tested applications | Sandwich ELISA |
| Gene ID (NCBI) | 350 |
| Sample Type | Average Range |
| Human serum | 96% 94%-99% |
| Cell lysate | 100% 98%-104% |
| Sample | n mean ( pg/mL ) SD CV% |
| 1 | 16 1,464.4 100.9 6.9 |
| 2 | 16 345.2 22.7 6.6 |
| 3 | 16 192.1 8.9 4.6 |
| Storage Instructions | All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions. |
Apolipoprotein H (ApoH), also known as beta2-glycoprotein I (B2-GPI), is a plasma glycoprotein, primarily synthesized in the liver. It has an important function in blood coagulation and clearance of apoptotic bodies from the circulation. APOH is found on several classes of lipoproteins, and is involved in the activation of lipoprotein lipase in lipid metabolism. ApoH is also an important actor of host innate immune response through its capacity to bind with high affinity to a large panel of pathogens or their proteins. Properties Storage Instructions All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions. Synonyms Apolipoprotein H, Activated protein C-binding protein, Apo H, Apo-H, ApolipoproteinH FAQ 1、I have limited sample diluent. How should I handle samples requiring high dilution factors? Answer: We recommend performing a serial dilution. For example, if human serum requires a 1,000,000-fold dilution, you can use a three-step 100× dilution method as follows: Step 1: Take 2 uL of human serum and add it to 198 uL of sample diluent, mix thoroughly to obtain a 100-fold diluted sample. Step 2: Take 2 uL of the 100-fold diluted sample and add it to 198 uL of sample diluent, mix thoroughly to obtain a 10,000-fold diluted sample. Step 3: Take 2 uL of the 10,000-fold diluted sample and add it to 198 uL of sample diluent, mix thoroughly to obtain a 1,000,000-fold diluted sample. If further dilution is required, additional serial steps can be used. We do not recommend using any other solutions or water for dilution, except for the sample diluent we suggest. If your sample diluent is still insufficient, please contact our customer service for more sample diluent.