| Product name | Human NCR1 ELISA Kit |
| Synonyms | CD335, hNKp46, LY94, Natural killer cell p46-related protein, NCR 1 |
| Reactivity | Human |
| Host / Isotype | United States |
| Tests | 1 X 96 well plate |
| Sample type | Serum, Plasma, Cell culture supernatant |
| Assay type | Sandwich |
| Sensitivity | 3.8 pg/mL |
| Range | 39-2500 pg/mL |
| Tested applications | Sandwich ELISA |
| Gene ID (NCBI) | 9437 |
| Sample Type | Average Range |
| Human serum | 84% 77%-94% |
| Cell culture supernatant | 102% 99%-107% |
| Sample | n mean ( pg/mL ) SD CV% |
| 1 | 16 1,232.1 60.9 4.9 |
| 2 | 16 307.9 15.3 5.0 |
| 3 | 16 164.6 9.0 5.4 |
| Storage Instructions | All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions. |
NKp46, also known as NCR1 or CD335, is an Ig-like superfamily cell surface receptor that is highly conserved in mammals. NKp46 is a type I transmembrane glycoprotein consisting of two extracellular Ig-like domains, a transmembrane domain, and an intracellular tail. It is expressed on NK cells, rare T-cell subsets, and a mucosal population of NKp46+ innate lymphoid cells. NKp46 is the major triggering receptor involved in natural cytotoxicity. Properties Storage Instructions All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions. Synonyms CD335, hNKp46, LY94, Natural killer cell p46-related protein, NCR 1 FAQ 1、I have limited sample diluent. How should I handle samples requiring high dilution factors? Answer: We recommend performing a serial dilution. For example, if human serum requires a 1,000,000-fold dilution, you can use a three-step 100× dilution method as follows: Step 1: Take 2 uL of human serum and add it to 198 uL of sample diluent, mix thoroughly to obtain a 100-fold diluted sample. Step 2: Take 2 uL of the 100-fold diluted sample and add it to 198 uL of sample diluent, mix thoroughly to obtain a 10,000-fold diluted sample. Step 3: Take 2 uL of the 10,000-fold diluted sample and add it to 198 uL of sample diluent, mix thoroughly to obtain a 1,000,000-fold diluted sample. If further dilution is required, additional serial steps can be used. We do not recommend using any other solutions or water for dilution, except for the sample diluent we suggest. If your sample diluent is still insufficient, please contact our customer service for more sample diluent.