T7 RNA Polymerase ELISA kit _ 36705ES yeasen

$495.00
In stock
SKU
36705ES48
Catalog No.SizePrice (USD)
36705ES4848 T$495.00
36705ES9696 T$850.00

Product Description

T7 RNA polymerase uses double-stranded DNA containing the T7 promoter sequence (5’-TAATACGACTCACTATAG*-3’) as a template and NTPs as substrates to synthesize RNA complementary to the reverse strand of DNA downstream of the promoter. Both linear double-stranded DNA with blunt ends or 5’ overhangs can serve as template substrates for T7 RNA polymerase, meaning linearized plasmids or PCR products can be used as templates for in vitro RNA synthesis. This kit employs a sandwich enzyme-linked immunosorbent assay (ELISA) principle to detect residual T7 RNA polymerase. T7 RNA polymerase standards and test samples are added to microplate wells pre-coated with anti-T7 RNA polymerase antibodies (36705-A). Diluted biotin-labeled T7 RNA polymerase detection antibodies (36705-C) are then added, followed by Streptavidin-HRP (SA-HRP) (36705-D), forming an antibody + antigen + antibody-biotin + SA-HRP complex. After washing, TMB substrate (36705-H) is added for color development. TMB is catalyzed by HRP to transition from colorless to blue and finally to yellow upon addition of stop solution (36705-I). The intensity of the yellow color is positively correlated with the amount of T7 RNA polymerase detected in the sample.

Validation Report

Features

  • High Sensitivity
  • : Quantitative lower limit (LOQ) of 1 ng/mL; detection limit (LOD) as low as 0.318 ng/mL.
  • High Precision
  • : Intra-assay CV <10%; intermediate precision (inter-assay CV) <15%.
  • High Accuracy
  • : Spike recovery rates consistently within 70–130%.
  • Strong Specificity
  • : No significant variation observed in T7 RNA polymerase values after adding various interfering substances.
  • Excellent Stability
  • : Stable for 7 days at 37 °C without performance loss.
  • Regulatory Compliance
  • : Fully validated in accordance with ChP, USP, and ICH Q2 (R2) guidelines.
  • Quality Assurance
  • : All critical raw materials, including coated and detection antibodies and standards, are independently developed and fully traceable.
  • Audit-Ready
  • : Stable manufacturing process with controlled lot-to-lot variation and comprehensive audit documentation.

Specifications

  • Parameter
  • Detection Range
  • 1 – 32 ng/mL
  • Assay Method
  • Double-antibody sandwich
  • Assay Time
  • 3.5 hours
  • Sensitivity
  • 0.318 ng/mL
  • Dilution Linearity
  • Difference <20% compared with undiluted samples
  • Recovery Rate
  • 70% – 130%
  • Intra-assay Variation
  • <10%
  • Inter-assay Variation
  • <15%

Components

  • Components No.
  • Name
  • 36705
  • S
  • 48
  • (48 T)
  • 36705
  • ES
  • 96
  • (96 T)
  • 36705-A
  • Anti-T7 RNA Polymerase coated microtiter strips
  • 48
  • T
  • 96 T
  • 36705-B*
  • Standard: T7 RNA Polymerase
  • 1
  • vial
  • 2
  • vial
  • 36705-C
  • Detection
  • Antibody
  • Biotin-conjugated
  • Antibodies
  • 60 μL
  • 120 μL
  • 36705-D
  • Streptavidin-HRP
  • 30 μL
  • 60 μL
  • 36705-E
  • Dilution Buffer 1
  • 25
  • mL
  • 45
  • mL
  • 36705-F
  • Wash Buffer Concentrate (20×)
  • 25
  • mL
  • 50
  • mL
  • 36705-G
  • Dilution Buffer 2
  • 15
  • mL
  • 30
  • mL
  • 36705-H
  • TMB Substrate
  • 8
  • mL
  • 15
  • mL
  • 36705-I
  • Stop Solution
  • 5
  • mL
  • 10
  • mL
  • 36705-J
  • Plate Sealer
  • 3 each
  • 5 each

Storage

Store at 2-8°C. The shelf life is 1 year when unopened, and 6 months after opening. Upon receipt of the product, please check if all components are complete, and immediately store them at the corresponding storage temperature.

Figures

1. Good Linearity Table 1. Linearity Range: 1~32 ng/mL,R2=0.999 ,CV≤5%。 Std Concn. ( ng/mL ) AVG ( ng/mL ) Recovery ( % ) CV% Std 1 32 31.965 99.9% 2.2% Std 2 16 16.070 100.5% 5.0% Std 3 8 7.812 97.7% 2.1% Std 4 4.192 104.8% 0.8% Std 5 2 2.020 101.8% 3.5% Std 6 1 1.173 117.3% 0.0% NC 0

Figure 1. Standard curve ) for .T7 RNA Polymerase ELISA kit 2. High Specificity 2 . Cross-reactivity comparison. Eight process-related proteins were tested using the Cat#36705ES kit to evaluate potential cross-reactivity of the T7 RNA Polymerase antibody with nine different proteins. Positive control (PC, 1 ng/mL) and negative control (NC, 0 ng/mL) were included. Results showed that the eight process-related proteins (10 µg/mL) produced signals comparable to the negative control, with detected concentrations below the kit’s quantitation limit (1 ng/mL), indicating no cross-reactivity. Table 2. Cross-reactivity Data with Proteins Added in Eight Manufacturing Processes Protein Type s OD Value Concentration ( ng/mL ) PC- 1 ng/mL 0.165 1 NC-0 ng/mL 0.123 0 UltraNuclease 0.125 n.d. Salt Active UltraNuclease 0.137 n.d. DNase I 0.161 n.d. Murine RNase Inhibitor 0.151 n.d. Vaccinia Capping Enzyme

0.128 n.d. Inorganic Pyrophosphatase 0.127 n.d. mRNA Cap 2´-O-Methyltransferase 0.125 n.d. 1%BSA 0.131 n.d. 3. High Accuracy: Spike recovery rates consistently within 70–130%. Table 3 . Dilution Linearity and Spike Recovery Sample Dilution Measured Value (ng/mL) Linearity ( % ) Sample + Spike Measured Value (ng/mL) Recovery (%) S + Std1 31.467 S + Std1 31.467 97.0% 2× 13.729 87.3% S + Std3 7.625 90.2% 4× 6.015 87.6% S + Std5 2.248 91.6% 8× 3.068 102.0% S (Sample) 0.907 16× 1.742 113.6% Equal volumes of high, medium, and low concentrations of T7 standards were spiked into mRNA purification samples, with the high-level spiked sample further subjected to serial dilutions. T7 detection results from diluted samples, spiked samples, and unspiked controls all showed linearity and spike recovery rates within 80%–120%. Spike Recovery Rate = Measured Value / (50% Spiked Standard + 50% Sample Value)

4. High Sensitivity: Quantitative lower limit (LOQ) of 1 ng/mL; detection limit (LOD) as low as 0.318 ng/mL. Table 4. LOQ Std6 theoretical value(ng/mL) Std 6 test mean value(ng/mL) Test repetitions CV% Recovery 1 0.858 24 12.6% 85.8% The limit of quantification for this product is determined by diluting the lowest concentration point of the standard curve, where the coefficient of variation (CV) at the lowest concentration is less than 20%. This is defined as the limit of quantification, which is 1 ng/mL. Table 4. LOD Mean value ( OD ) SD ( OD ) Mean value +2SD (OD) Mean value +2SD(ng/mL) 0.0527 0.0038 0.0604 0.3176 The detection limit of this product is defined as the standard concentration corresponding to the mean detection value of the negative control (NC) plus two times the standard deviation. By measuring 24 NC standards, calculating the mean and standard deviation, the detection limit is determined to be 0.318 ng/mL.

Safety Data Sheet

  • 36705_MSDS_HB250925.PDF

Manuals

  • 36705_Manual_Ver.EN20250925

Reviews

Write Your Own Review
You're reviewing:T7 RNA Polymerase ELISA kit _ 36705ES yeasen
Your Rating
Copyright © 2025 Biogege, Inc. All rights reserved.