CHO Host Cell DNA Residue Detection Kit (3G) _ 41332ES yeasen

$1,155.00
In stock
SKU
41332ES50
Catalog No.SizePrice (USD)
41332ES5050 T$1,155.00
41332ES60100 T$2,075.00

Product Description

CHO Host Cell DNA Residue Detection Kit is used for the quantitative analysis of CHO host cell DNA residuce in intermediate samples, semi-finished and finished products of various biological products. This kit adopts Taqman fluorescent probe and the polymerase chain reaction (PCR) method, which has fg level minimum detection limit and can specifically and quickly detect the residual CHO cell DNA. The kit needs to be used together with the the Residual DNA Sample Preparation Kit (Cat# 18461ES or 18469).

Features

  • High anti-interference capability
  • – Reliable performance in complex sample matrices; spike recovery rates consistently between 70–130%.
  • High sensitivity
  • – Lower limit of quantification (LLOQ) as low as 0.3 fg/μL.
  • Excellent precision
  • – Intra-assay CV <10%; inter-assay (intermediate precision) CV <15%.
  • Strong specificity
  • – Exclusively detects CHO residual DNA without interference from other genomic DNA.
  • Comprehensive validation
  • – Fully validated according to ICH Q2 (R2) guidelines; complete validation reports available.
  • Regulatory compliance
  • – Verified in accordance with EP 2.6.7, JP G3, and USP <63> pharmacopeia requirements, meeting international standards.
  • Quality assurance
  • – All enzyme raw materials are self-developed and industrialized, ensuring stable supply and consistent quality.

Components

  • Components No.
  • Name
  • 41332ES50
  • (50 T)
  • 41332ES60
  • (100 T)
  • A
  • CHO qPCR Mix
  • 0.75 mL
  • 1.5 mL
  • B
  • CHO Primer&Probe Mix
  • 250 μL
  • 500 μL
  • C
  • DNA Dilution Buffer
  • 2×1.8 mL
  • 4×1.8 mL
  • D
  • CHO DNA Control(30 ng/μL)
  • 25 μL
  • 50 μL

Specifications

Standard curve range3 fg/μL – 300 pg/μL
Correlation coefficient (R²)≥ 0.98
Slope–3.1 to –3.8
Amplification efficiency90%–110%
Recovery rate50%–150% (industry standard 70%–130%)
RepeatabilityCV < 15%
Intermediate precisionCV < 15%
SensitivityLimit of quantification (LOQ): 0.5 fg/μL
SpecificityNo interference from other genomic DNA sources Stability Accelerated stability: 14 days at 37 °C Freeze–thaw stability: ≥ 10 cycles Blank Control No-template control (NTC): In 96 repeated qPCR runs, all Ct values ≥ 35 or no amplification observed Applicable instrument models Bio-Rad: CFX96 Optic Module. Thermo Scientific: ABI 7500; ABI Quant Studio 5. SLAN-96S.

Applications

  • Detecting the residual DNA content of CHO host cells in products such as antibodies, recombinant proteins, and vaccines produced using CHO cells.

Storage

This product should be stored at -25~-15℃ for 1 year.

Figures

1. Good Linearity

Figure 1. Standard curve (left) and amplification curve plot (right) for CHO DNA (3G). The linear range of the CHO Host Cell Residual DNA Detection Kit (3G) is 3 fg/μL to 300 pg/μL, with an R² value of 1.0 and an amplification efficiency of 99.29%. The coefficient of variation (CV) for detection values across concentrations is <15%. 2. Sensitivity: Limit of quantification (LOQ) - 0.3 fg/μL

Figure 2. qPCR detection of CHO DNA at 0.3 fg/μL. CHO DNA (3G) at the lowest standard curve concentration (3 fg/μL) and below was tested with 10 replicates per concentration. Results showed that at concentrations ≥0.3 fg/μL, the CV was <20%, indicating that the limit of quantification (LLOQ) of the CHO Host Cell DNA Residual Detection Kit (3G) is 0.3 fg/μL. 3. High Recovery rate: 50%–150% Table 1. Recovery of CHO DNA in Blank Samples Spike recovery rates for all concentrations tested were between 70% and 130%, with coefficients of variation (CV) below 20%, demonstrating consistent performance of the assay across a wide dynamic range. Table 2. CHO DNA Recovery in Various Sample Matrices CHO DNA (300 fg/μL) was spiked into five different background matrices (high protein, high nucleic acid, high salt, high pH, low pH) and extracted using the same procedure. Recovery rates ranged from 70% to 130%, with CVs below 20% for all matrices, demonstrating robust performance across diverse sample backgrounds.

Safety Data Sheet

  • 41332_MSDS_HB250924_EN.PDF

Manuals

  • 41332_Manual_Ver.EN20240517.PDF

Related Blog

  • Host Cell Residual DNA Detection Kits facilitate Biologics R&D and Manufacturing

Reviews

Write Your Own Review
You're reviewing:CHO Host Cell DNA Residue Detection Kit (3G) _ 41332ES yeasen
Your Rating
Copyright © 2025 Biogege, Inc. All rights reserved.