Product Description
CHO Host Cell DNA Residue Detection Kit is used for the quantitative analysis of CHO host cell DNA residuce in intermediate samples, semi-finished and finished products of various biological products. This kit adopts Taqman fluorescent probe and the polymerase chain reaction (PCR) method, which has fg level minimum detection limit and can specifically and quickly detect the residual CHO cell DNA. The kit needs to be used together with the the Residual DNA Sample Preparation Kit (Cat# 18461ES or 18469).
Features
Components
Specifications
| Standard curve range | 30 fg/μL – 300 pg/μL |
| Correlation coefficient (R²) | ≥ 0.98 |
| Amplification efficiency | 90%–110% |
| Recovery rate | 50%–150% (industry standard 70%–130%) |
| Repeatability | CV < 15% |
| Intermediate precision | CV < 15% |
| Sensitivity | Limit of quantification (LOQ): 10 fg/μL |
| Specificity | No interference from other genomic DNA sources Stability Accelerated stability: 14 days at 37 °C Freeze–thaw stability: ≥ 10 cycles Blank Control No-template control (NTC): In 96 repeated qPCR runs, all Ct values ≥ 35 or no amplification observed Applicable instrument models Bio-Rad: CFX96 Optic Module. Thermo Scientific: ABI 7500; ABI Quant Studio 5. SLAN-96S. |
Applications
Storage
This product should be stored at -25~-15℃ for 2 years. Both 41331-A and 41331-B should be stored protected from light.
Figures
1. Good Linearity
Figure 1. Standard curve (left) and amplification curve plot (right) for HEK293 DNA (3G). The linear range of the HEK293 Host Cell Residual DNA Detection Kit (3G) is 30 fg/μL to 300 pg/μL, with an R² value of 1.0 and an amplification efficiency of 100.45%. The coefficient of variation (CV) for detection values across concentrations is <15%. 2. Sensitivity: Limit of quantification (LOQ) - 0.3 fg/μL
Figure 2. qPCR detection of HEK293 DNA at 10 fg/μL. HEK293 DNA (3G) at the lowest standard curve concentration (3 fg/μL) and below was tested with 10 replicates per concentration. Results showed that at concentrations ≥10 fg/μL, the CV was <20%, indicating that the limit of quantification (LOQ) of the HEK293 Host Cell DNA Residual Detection Kit (3G) is 10 fg/μL. 3. High Recovery rate: 50%–150% Table 1. Recovery of HEK293 DNA in Blank Samples Spike recovery rates for all concentrations tested were between 70% and 130%, with coefficients of variation (CV) below 20%, demonstrating consistent performance of the assay across a wide dynamic range. Table 2. HEK293 DNA Recovery in Various Sample Matrices HEK293 DNA (3 pg/μL) was spiked into five different background matrices (high protein, high nucleic acid, high salt, high pH, low pH) and extracted using the same procedure. Recovery rates ranged from 70% to 130%, with CVs below 20% for all matrices, demonstrating robust performance across diverse sample backgrounds.
Safety Data Sheet